No. 6
Aerosol Photometer Recalibration

Frequently, customers inquire about recalibrating ATI Aerosol Photometers. As a manufacturer of aerosol photometers, ATI can only make recommendations about recalibration. The actual recalibration interval is controlled by each individual's corporate or company quality assurance program. For example, the Charleston Naval Shipyard has a Quality Assurance Program which dictates that any instrument using an analog meter indicator shall be recalibrated on a semi-annual basis. Therefore, they return the instrument every six months for recalibration to comply with their internal Quality Assurance Program.

ATI has numerous validated calibration records which demonstrate that the current TDA-2D, 2E & 2G Series Photometers maintain their accuracy for a minimum of one year under normal usage. Therefore, ATI's recommended maximum photometer recalibration interval is annually.