|

No. 15
Leak Testing with Particle Counters
Problems Testing Filter Systems with a Particle
Counter
- Particle counters count the number of particles in a
cubic foot of air. To do so accurately, a sample must
be taken at a specific location for 1 minute. If a sample
is taken for only 6 seconds (one-tenth of a minute), a
'0' must be added to the total count, introducing a possible
10 % error.
- Particle counters will only give accurate counts up
to approximately 300,000 particles per cubic foot per
minute (or 30,000 particles per cubic foot in 6 seconds).
Concentrations higher than this introduce large COINCIDENCE
LOSS which leads to errors of 5-20%, depending on the
aerosol concentration and the particle counter. It is
a acknowledged fact that two identical particle counters
manufactured and calibrated by the same company will not
produce identical results when exposed to the same aerosol
sample.
- A good example of the accumulation of errors is the
Palas' filtration system, which purports to use 106 to
107 particles per cubic centimeter which requires diluters
to sample the upstream challenge concentration. This equates
to 2.83x10E10 to 2.83x10E11 particles per cubic foot.
The diluters are 10:1 units which have a 5-10% error margin.
By placing 2 diluters in series, the error margin increases
to 10-20%, and gives a dilution of 100:1. The lower concentration
is 2.83x10E10 particles/cu. ft. = 2.83x10E10/100 = 2.83x108
particles/cu. ft. or 283 million particles per cubic foot.
Particle Counters CANNOT count that many particles accurately.
Therefore, any downstream sample reading, even for less
than 1 minute, would not be reliable.
- From a practical standpoint (forgetting diluter error,
count error, etc.) and just calculating scanning time,
the Palas filtration test system becomes unwieldy. Using
an Isokinetic Probe, per IEST Standards, sampling at 1
cfm and with an opening of 3" x 0.5", it would take [384
readings (counts) at 6 seconds each] 38.4 hours to scan
a 24" x 24" HEPA filter. Remember, in addition, a 10 -
20% error rate due to extrapolating what the particle
count would be for a full minute by multiplying the count
obtained in just 6 seconds by 10.
A particle counter counts the number of particles
passing through its viewing area. Most particle counters don't
pass the entire 1 cfm sample through the viewing area but
only a portion which introduces more room for error. Knowing
the sample flow rate (1 cfm) and the duration of sample time
(1 minute), it can be determined how many particles were in
that one cubic foot sampled for one minute. This is what they
were designed to do, and as such, make an excellent monitor
for clean or sterile areas.
An aerosol photometer is a true ratio detector
since it measures particles "en masse" and all of the particles
travel through its viewing or detection area. ATI photometers
are capable of measuring concentrations of particles as high
as 600 micrograms per liter and are sensitive enough to measure
accurately and instantaneously concentrations down to 0.00001
micrograms per liter. Using an ATI photometer, the same 24"
x 24" HEPA filter can be scanned in accordance with IEST and
other Standards in 1.6 minutes. If there is any area that
has leakage > .01%, it can be pinpointed in seconds.
An excellent article on this subject was
printed in Performance Review, a peer reviewed technical journal
published quarterly by the Controlled Environment Testing
Association (CETA). The article titled HEPA & ULPA Filter
Installation Leak Tests by Ulrich Dietrich appeared in the
summer of 1995 edition. For a copy, contact CETA at (202)
737-0204.
|