|

1) What is the required upstream aerosol level
for leakage testing?
2) Will using aerosol to test filters cause
excessive loading (clogging) and shorten the filters usable
life?
3) Where should upstream challenge aerosol
be introduced into a filter system that is being tested?
4) What type of PAO does the FDA require for
filter leakage certification tests?
5) How much compressed air does a Type III-A
Laskin nozzle aerosol generator need?
6) What is the particle size distribution of
a Laskin nozzle generator using PAO?
7) What is the particle size distribution
of a Laskin nozzle generator using DOP?
8 ) What is the particle size distribution of a Thermal generator
(TDA-5A/5B) using PAO?
9 ) What is the particle size distribution
of a Thermal generator (TDA-5A/5B) using DOP?
10) What is the recommended shelf life of PAO
(CAS # 68649-12-7)?
11) Can a TDA-5B be used to test Biological
Safety Cabinets (BSC's)?
12) What happens if the photometer sample flow drops below its normal operating range of 28.3 lpm +/- 2.8?
13) What is the effect of the "Straylight" value on photometer operation?
1) What is the required
upstream aerosol level for leakage testing?
Most current standards and recommended practices require
a minimum of 10 micrograms per liter (ug/l) as an upstream
aerosol challenge. While higher concentrations can be used,
10-ug/l is the minimum that may be used and still have a valid
leakage certification test. All of ATI's aerosol photometers
were designed to operate at upstream challenge concentrations
between 10 and 100 ug/l.

2) Will using aerosol
to test filters cause excessive loading (clogging) and shorten
the filters usable life?
Poly-dispersed aerosol is used to challenge the "integrity"
or leakage of the plenums, framework, connections, etc., and
the filter media itself. Therefore people ask, "How much aerosol
is needed?"
The example below uses 50 micrograms per liter as an upstream
challenge.
Example: Let's use a filter bank that is 10 feet x 10 feet
and utilizes (25) 24" x 24" x 12" HEPA filters. The normal
24" x 24" x 12" HEPA has an average of 65 pleats. The pleat
size is 22 1/2" x 10 1/2". If we take this information and
multiply it, we come up with the average HEPA filter containing
approximately 213 square feet of media. If you multiply this
by 25 (for the 25 HEPA filters in the bank) we come up with
a total of 5,325 square feet of media absorb. We will also
calculate the amount of liquid we will aerosolize and let
this 5,325 square feet of media absorb. To figure out how
much liquid we will introduce to these filters, we have to
calculate how long it will take to scan the filter bank. If
you use a photometer with a rectangular isokinetic probe,
it will take 40 minutes. This is calculated by figuring the
scanning rate of 10 feet per minute. Since the filter bank
is 10 feet wide, one scan per minute across the face of the
filters would be allowed. Since the probe dimension is 3 inches
and the bank is 120 inches high, 3 into 120 equals 40 strokes.
The test time will be 40 minutes (each stroke taking 1 minute,
it requires 40 strokes to cover the entire area).
The next calculation is the amount of liquid that will be
converted into an aerosol and spread over the 5,325 square
feet of media. This equals 0.00283 grams per cfm (calculate
100 micrograms per liter x 28.3 for one cubic foot per minute).
Since the system is a 25,000 cfm system, we multiply 25,000
x .00283 grams per cfm which totals 70.75 grams of liquid
we will aerosolize to obtain a challenge of 100 micrograms
per liter in this 25,000 cfm filtration system. Because we
will run this test for 40 minutes, we will multiply 70.75
x 40 minutes and come up with a total of 2,830 grams of liquid
that we will aerosolize to challenge the in-tegrity of the
25,000 cfm filtration system with 100 micrograms per liter.
The 2,830 grams is equal to 99.8 ounces of oil spread over
5,325 square feet of media (equivalent to a 12 inch wide roll
of filter media just under a mile long). How can we clog this
filter system? If we decide to only use 36 ounces to test
the filter system, we could then challenge it with 36.1 micrograms
per liter (equal to three-12 ounce containers).

3) Where should
upstream challenge aerosol be introduced into a filter system
that is being tested?
Adequate aerosol mixing upstream can usually be obtained by
introduction at least ten duct diameters upstream of the filters,
or by introducing it upstream of baffles or turning vanes
in the duct. When neither of these methods is practical, a
Stairmand disk located four to six duct diameters upstream
will provide satisfactory mixing. A Stairmand disk is a plate
with the same geometric shape as the duct section that blocks
the central half of the duct area. Air flowing past the disk
creates vortices on the leeward side that compel turbulent
and thorough mixing of the introduced aerosol and the dilution
air stream. The required diameter of the stairmand disk
may be calculated using the following information.
The Stairmand disk diameter equals the pipe (or duct) diameter
divided by the
square root of two (1.414213).

4) What type
of PAO does the FDA require for filter leakage certification
tests?
Several differing opinions on the correct PAO to use have
arisen recently. The FDA, in their original CGMP (Current
Good Manufacturing Processes) release was specific in the
type that was approved as a replacement for DOP (DEHP).
"The
Chemical Abstracts Service (CAS) number which identifies this
product also remained as 68649-12-7. Other reported alternatives
used in the industry include DOS (Di-(2-ethylhexyl) sebacate)
and Ondina Oil. However, no manufacturer has yet submitted
all the necessary data to evaluate these alternatives. As
such, Emery 3004 PAO with the CAS number 68649-12-7 still
remains an acceptable replacement for DOP."
CGMP
Notes (Vol. 4, Number 4), December 1996.
ATI's position is that PAO with a CAS # 68649-12-7 is acceptable
by FDA definition. The full text of the referenced CGMP document
is available at: http://www.fda.gov/cder/hdn/cnotesd6.htm.

5) How much
compressed air does a Type III-A Laskin nozzle aerosol generator
need?
Each Laskin nozzle being used in a generator consumes approximately
2.64 cfm (75 Liters) of air at 20 psi (1.4 bars) and total
air consumption is proportional to the number of nozzles in
use. Therefore the maximum compressed air requirement for
a three nozzle generator is 7.92 cfm at 20 psi and a six nozzle
generator would need 15.84 cfm at 20 psi. The applied pressure
of 20 psi needs to remain constant to allow calculation of
the aerosol generator output. Air compressor performance specifications
that will show output volume (cfm) versus pressure (psi) are
readily available from most vendors and will make selection
of an appropriate compressor less difficult.

6) What is the
particle size distribution of a Laskin nozzle generator using
PAO?
TDA-4Blite (Type III-A Laskin nozzle) @ 20 psi using PAO (Emery
3004)
|
|
Number |
Surface |
Mass |
Volume |
| Particle Size |
Particle Size |
Particle Size |
Particle Size |
| 245 |
415 |
528 |
528 |
| 281 |
454 |
549 |
549 |
| 248 |
407 |
503 |
503 |
| 233 |
429 |
594 |
594 |
| 1.65 |
1.62 |
1.55 |
1.55 |
|

7)
What is the particle size distribution of a Laskin nozzle
generator using DOP?
TDA-4Blite (Type III-A Laskin nozzle) @ 20 psi
using DOP (DEHP)
|
Number |
Surface |
Mass |
Volume |
| Particle Size |
Particle Size |
Particle Size |
Particle Size |
| 254 |
430 |
546 |
546 |
| 291 |
468 |
563 |
563 |
| 256 |
420 |
517 |
517 |
| 241 |
429 |
685 |
685 |
| 1.66 |
1.61 |
1.54 |
1.54 |
|
8) What is
the PAO particle size distribution of a Thermal generator
(TDA-5A/5B)?
TDA-5B Thermal Condensation Type Aerosol Generator using
PAO (Poly-alpha Olefin)

|
Number |
Surface |
Mass |
Volume |
| Particle Size |
Particle Size |
Particle Size |
Particle Size |
| 257 |
321 |
356 |
356 |
| 273 |
334 |
365 |
365 |
| 259 |
318 |
350 |
350 |
| 241 |
334 |
372 |
372 |
| 1.41 |
1.37 |
1.35 |
1.35 |
*The TDA-5B aerosol distribution listed above is characteristic of the operating conditions and settings present at the time of testing. Particle size distributions generated during field usage will change depending upon the ambient temperature, humidity and equipment settings in use.

9)
What is the DOP/DEHS particle size distribution of a Thermal
generator (TDA-5A/5B)?
TDA-5B Thermal Condensation Type Aerosol Generator
using DOP (Dioctyl Phythalate)
|
Number |
Surface |
Mass |
Volume |
| Particle Size |
Particle Size |
Particle Size |
Particle Size |
287 |
354 |
384 |
384 |
296 |
361 |
392 |
392 |
275 |
346 |
377 |
377 |
300 |
385 |
400 |
400 |
1.54 |
1.35 |
1.33 |
1.33 |
*The TDA-5B aerosol distribution listed above is characteristic of the operating conditions and settings present at the time of testing. Particle size distributions generated during field usage will change depending upon the ambient temperature, humidity and equipment settings in use.
10)
What is the recommended shelf life of PAO (CAS # 68649-12-7)?
The manufacturer's stated shelf life for PAO (CAS# 68649-12-7) is three (3) years
in the original, sealed and unopened container. The
shelf life of an open container, that is tightly sealed between
uses, is one (1) year.

11)
Can a TDA-5B be used to test Biological Safety Cabinets (BSC's)?
ATI does not recommend the use of the TDA-5B or any other
thermal condensation based aerosol generator for use in the
certification of Biological Safety Cabinets. IEST RP034
specifically cautions against this practice and NSF/ANSI 49-2002,
Annex F specifcally calls out for the use of a Laskin nozzle
style generator or equivalent during certification.
A TDA-5B generator is not considered a Laskin nozzle equivalent
aerosol generator.

12) What happens if the photometer sample flow drops below its normal operating range of 28.3 lpm +/- 2.8?
Reduction of the photometer sample flow to a level below the lower tolerance limit, while using the internal reference settings for the 100% level, results in an increased displayed leakage value. This higher value is the result of an increase in residence time of the aerosol passing through the light scattering chamber's area of observation. The longer residence time of the aerosol results in a more conservative evaluation of the filter under test. If an actual upstream aerosol sample was being used to establish a 100% point, the unit’s aerosol response is self correcting for the decreased flow rate and standard accuracy conditions would be maintained.

13) What is the effect of the "Straylight" value on photometer operation?
The straylight value is an indication of the optical conditions of the Light Scattering Chamber (LSC) at a given point in time and as such, do not have a tolerance range established. Increases in the straylight usually occur through ordinary usage over the annual calibration cycle. Sudden and/or catastrophic increases in the straylight value typically are the result of operator error or component failure and render the unit inoperable. Some examples of catastrophic failure are introduction of liquid to the scattering chamber, formation of condensation on the optics, large accumulations of solid particulate and breakage of discrete components within the LSC. Photometers which remain operable, even with elevated straylight values, are still capable of accurately measuring filter leakage values.

|